3M™ Attest™ eBowie-Dick Test Card, 10135, 30/BAG, 1 BAG/CS, 30/CS

Hover over image to zoom

3M™ Attest™ eBowie-Dick Test Card, 10135, 30/BAG, 1 BAG/CS, 30/CS

About the product

COMING SOON - 3M™ Attest™ eBowie-Dick Test Card is designed to monitor equipment performance of steam sterilizers. It is a single-use test card and provides a "pass" or "fail" test result in less than 5 seconds when inserted into the 3M™ Attest™ eBowie-Dick Auto-reader.

Product Details

See the components you need for the 3M™ Attest™ eBowie-Dick Test System.
3M™ Attest™ eBowie-Dick Auto-reader 1190

3M™ Attest™ eBowie-Dick Auto-reader 1190

3M™ Attest™ eBowie-Dick Test Card 10135, 10135B

3M™ Attest™ eBowie-Dick Test Card 10135, 10135B

3M™ Attest™ eBowie-Dick Card Holder 10135CH, 10135CHB

3M™ Attest™ eBowie-Dick Card Holder 10135CH, 10135CHB

3M™ Attest™ eBowie-Dick module within 3M™ Attest™ Connect PC Application

3M™ Attest™ eBowie-Dick module within 3M™ Attest™ Connect PC Application

  • 3M™ Attest™ eBowie-Dick Test System eliminates the guesswork of determining pass/fail results.
  • Meets ISO 11140 5:2007/(R)2012 and is a Type 2 (Category s2) Special Indicator as categorized by ISO 11140 1:2014.
  • Is equivalent in performance to the Bowie-Dick towel pack described in AAMI ST-79.
  • Contains a process chemical indicator on top left of card.
  • Card must be placed into 3M™ Attest™ eBowie-Dick Auto-reader for final test results.

Suggested Applications

  • The 3M™ Attest™ eBowie-Dick Test Card 10135 and 3M™ Attest eBowie-Dick Card Holder 1015CH are used in conjunction with the 3M™ Attest™ eBowieDick Auto-reader 1140 to monitor the equipment performance of steam sterilizers each day, as well as after a steam sterilizer repair
  • The 3M™ Attest™ eBowie-Dick Test Card 10135 will detect steam penetration as a means of measuring air removal efficiency
  • The 3M™ Attest™ eBowie-Dick Test Card 10135 and 3M™ Attest™ eBowie-Dick Card Holder 10135CH is designed for testing air remo8al efficiency of 132°C (270°F), 134°C (273°F), and 135°C (275°F) for 3.5 minute test cycles in dynamic‑air‑remo8al steam sterilizers with chamber volumes greater than 54L (approximately 2 cubic feet)

Product Specifications

Brand
Brand

Attest™

Sterilization